Related Point of Care Diagnostic Devices





 |
 |
Hemo Control Blood Analyzer
for the fast and precise determination of hemoglobin and hematocrit in blood
|
|
|
|
| overview |
features |
technical |
accessories |
downloads |
The Hemo_Control system enables the fast and precise determination of hemoglobin and hematocrit in blood. It consists of a handy photometer, operating with microcuvettes (dry chemistry). Power supply by internal, rechargeable batteries or regular mains (100-250 V, 50/60 Hz).

After the uptake of a small venous, arterial or capillary blood sample using microcuvette it has to be inserted into the Hemo_Control photometer. Within a few seconds the current level of hemoglobin and hematocrit is displayed, added by other parameters, limitations and patient information on demand.
The Hemo_Control system is totally easy to use, highly reliable and well established at very satisfied customers around the world.
- Hemoglobin and hematocrit from one sample
- Determination in approx. 25 sec.
- Only 8 µL sample volume required
- Wide measurement range
- Comfort touch screen
- Maintenance-free
- Cost-efficient cuvettes
- Easy and flexible handling
- Ready for mobile and stationary use
- No additional reagents or dilutions required
Technical Data - Hemo_Control blood analyzer
| 0 – 25,6 g/dL |
| 8 µL |
| Venous, arterial or capillary human blood |
| 0-20 g/dL ± 0,3 g/dL; >20 g/dL ± 0,7 g/dL |
| ≤ 2 % |
| Approx. 25 sec. |
| Up to 4,000 results including date and time, patient type |
| In-plant according to NCCLS standard |
| Control cuvette (included), 3 different control solutions, Hb-con low, Hb-con norm, Hb-con high |
| Hemoglobin microcuvette |
| Photometric, azidemethemoglobin |
| Automatic |
| Power supply, 100 - 250 VAC/50 - 60 Hz, Integrated battery (approx. 100 h in use) |
| RS 232 C (Printer), PC (EDP cable) |
| 160 x 160 x 68 mm |
| approx. 700g |
| Humidity of < 90 % is advisable for the use of the device |
| +15 °C to +40 °C |
| 0 °C to +50 °C |
| +15 °C to +30 °C |
 |
 |
 |
box of 50 pcs. |
 |
depth of penetration: 2.2 mm, gauge: G22,
blood flow: medium 20...40 μL, SLN240, box of 100 pcs. |
 |
dropper bottle of 1 mL |
 |
2 dropper bottles of 1 mL with:
1 x Hb con-norm,
1 x Hb con-high |
 |
for cleaning the optical unit,
1 pc. or set of 5 pcs. |
 |
|
 |
with serial and
parallel interfaces |
| Hemo Control flyer |
1.3 MB |
 |
| Hemo Control Cleaning |
594 KB |
 |
| Hemo Control Sampling |
511 KB |
 |
| Hemo Control instructions |
137 KB |
 |
| Hemo Control haemoglobin |
902 KB |
 |
These are PDF files, to obtain a free Adobe PDF reader:  |
|